ATELLICA IM B·R·A·H·M·S PCT assay is for in vitro diagnostic use in the quantitative determination of procalcitonin (PCT) in human serum and plasma using the Atellica IM Analyzer.
ATELLICA IM B·R·A·H·M·S PCT is CE marked and FDA 510(k) cleared.
Assay characteristics
ATELLICA IM B·R·A·H·M·S PCT assay can be distinguished by these characteristics:
Test principle | One-step chemiluminescence immunoassay using three monoclonal antibodies (sandwich principle) |
Sample matrix | Serum, EDTA plasma, lithium heparin plasma, sodium heparin plasma |
Sample volume | 100 μL |
Time to result | 18 min |
Measuring range | 0.04-50.00 μg/L |
Functional sensitivity (LoQ) | 0.04 μg/L |
Analytical sensitivity (LoD) | 0.03 μg/L |