PCT-Collage LumipulseChemiluminescent enzyme immunoassay (CLEIA) for the determination of Procalcitonin (PCT) in human serum and plasma on the LUMIPULSE® G system (Fujirebio Inc.)

Assay characteristics

  • Two-step chemiluminescent enzyme immunoassay (sandwich principle)
  • Sample matrix: Serum, Plasma
  • Sample volume: 60 μL
  • Time to result: 30 min
  • Measuring range: 0.02 -100 μg/L
  • Functional sensitivity: 0.0079 μg/L
  • Analytical sensitivity: 0.0048 μg/L

Correlation to other PCT assays

Correlation between Lumipulse® G B·R·A·H·M·S PCT (Cartridge) and Thermo Scientific™ B·R·A·H·M·S PCT™ sensitive KRYPTOR™

Correlation Lumipulse kryptor

 

Precision

The Lumipulse® G B·R·A·H·M·S PCT demonstrated precision 4% (total %CV) in a study run according to the Clinical and Laboratory Standards Institute (CLSI) Protocol EP5-A2.

Precision Lumipulse

 

 